Measurement of spatial resolution of the micro-CT system

Get Citation
RONG Jun-Yan, FU Guo-Tao, WEI Cun-Feng, SUN Cui-Li and WEI Long. Measurement of spatial resolution of the micro-CT system[J]. Chinese Physics C, 2010, 34(3): 412-416. doi: 10.1088/1674-1137/34/3/022
RONG Jun-Yan, FU Guo-Tao, WEI Cun-Feng, SUN Cui-Li and WEI Long. Measurement of spatial resolution of the micro-CT system[J]. Chinese Physics C, 2010, 34(3): 412-416.  doi: 10.1088/1674-1137/34/3/022 shu
Milestone
Received: 2009-03-30
Revised: 2009-04-29
Article Metric

Article Views(2316)
PDF Downloads(623)
Cited by(0)
Policy on re-use
To reuse of subscription content published by CPC, the users need to request permission from CPC, unless the content was published under an Open Access license which automatically permits that type of reuse.
通讯作者: 陈斌, bchen63@163.com
  • 1. 

    沈阳化工大学材料科学与工程学院 沈阳 110142

  1. 本站搜索
  2. 百度学术搜索
  3. 万方数据库搜索
  4. CNKI搜索

Email This Article

Title:
Email:

Measurement of spatial resolution of the micro-CT system

    Corresponding author: WEI Long,

Abstract: 

The performance test is an important and necessary work for the micro-CT (computed tomography) system. The focal spot size of the micro focus X-ray tube is measured. The method of measuring the spatial resolution of micro-CT is introduced. A line-pair resolution of 28.2 lp/mm at the 10% modulation transfer function (MTF) level can be achieved with 14.7 μm spot size, 12.3 μm voxel size and a 25 mm field of view. In addition, a tungsten wire with the diameter of 5 μm can be detected by the system.