Geometric calibration for a SPECT systemdedicated to breast imaging

Get Citation
WU Li-Wei, CAO Xue-Xiang, WANG Lu, HUANG Xian-Chao, CHAI Pei, YUN Ming-Kai, ZHANG Yu-Bao, ZHANG Long, SHAN Bao-Ci and WEI Long. Geometric calibration for a SPECT systemdedicated to breast imaging[J]. Chinese Physics C, 2012, 36(10): 1019-1024. doi: 10.1088/1674-1137/36/10/019
WU Li-Wei, CAO Xue-Xiang, WANG Lu, HUANG Xian-Chao, CHAI Pei, YUN Ming-Kai, ZHANG Yu-Bao, ZHANG Long, SHAN Bao-Ci and WEI Long. Geometric calibration for a SPECT systemdedicated to breast imaging[J]. Chinese Physics C, 2012, 36(10): 1019-1024.  doi: 10.1088/1674-1137/36/10/019 shu
Milestone
Received: 2012-01-17
Revised: 2012-02-09
Article Metric

Article Views(1723)
PDF Downloads(288)
Cited by(0)
Policy on re-use
To reuse of subscription content published by CPC, the users need to request permission from CPC, unless the content was published under an Open Access license which automatically permits that type of reuse.
通讯作者: 陈斌, bchen63@163.com
  • 1. 

    沈阳化工大学材料科学与工程学院 沈阳 110142

  1. 本站搜索
  2. 百度学术搜索
  3. 万方数据库搜索
  4. CNKI搜索

Email This Article

Title:
Email:

Geometric calibration for a SPECT systemdedicated to breast imaging

    Corresponding author: SHAN Bao-Ci,

Abstract: Geometric calibration is critical to the accurate SPECT reconstruction. In this paper, a geometric calibration method was developed for a dedicated breast SPECT system with a tilted parallel beam (TPB) orbit. The acquisition geometry of the breast SPECT was firstly characterized. And then its projection model was established based on the acquisition geometry. Finally, the calibration results were obtained using a nonlinear optimization method that fitted the measured projections to the model. Monte Carlo data of the breast SPECT were used to verify the calibration method. Simulation results showed that the geometric parameters with reasonable accuracy could be obtained by the proposed method.