THE PERFORMANCE OF A LARGE AREA PLASTIC SCINTILLATION COUNTER FOR TOF MEASUREMENT

Get Citation
WANG Chao-Jun, WANG Dian-Rong, LIN Jian-Feng, JIANG Ying-Yuan, XU Yu-Lin and XI Ji-Wei. THE PERFORMANCE OF A LARGE AREA PLASTIC SCINTILLATION COUNTER FOR TOF MEASUREMENT[J]. Chinese Physics C, 1985, 9(4): 498-501.
WANG Chao-Jun, WANG Dian-Rong, LIN Jian-Feng, JIANG Ying-Yuan, XU Yu-Lin and XI Ji-Wei. THE PERFORMANCE OF A LARGE AREA PLASTIC SCINTILLATION COUNTER FOR TOF MEASUREMENT[J]. Chinese Physics C, 1985, 9(4): 498-501. shu
Milestone
Received: 1900-01-01
Revised: 1900-01-01
Article Metric

Article Views(2353)
PDF Downloads(463)
Cited by(0)
Policy on re-use
To reuse of subscription content published by CPC, the users need to request permission from CPC, unless the content was published under an Open Access license which automatically permits that type of reuse.
通讯作者: 陈斌, bchen63@163.com
  • 1. 

    沈阳化工大学材料科学与工程学院 沈阳 110142

  1. 本站搜索
  2. 百度学术搜索
  3. 万方数据库搜索
  4. CNKI搜索

Email This Article

Title:
Email:

THE PERFORMANCE OF A LARGE AREA PLASTIC SCINTILLATION COUNTER FOR TOF MEASUREMENT

    Corresponding author: WANG Chao-Jun,

Abstract: A large area plastic scintillation counter(100×20×1cm3) for TOF measurement has been made and the performance of this counter has been measured. A self researched new type Mean-Timer was used to much better eliminate the time difference of the fluo-rescence light transmission in the scintillator generated by the different location of the injecting particles, obtained time resolution of scintillation counter is about 347 ps.