A pilot experiment for mass measurement at CSRe
- Received Date: 2008-10-27
- Accepted Date: 2009-01-12
- Available Online: 2009-07-05
Abstract: A pilot experiment of mass measurement was performed at CSRe with the method of isochronous mass spectrometry. The secondary fragments produced via RIBLL2 with the primary beam of 400 MeV/u 36Ar delivered by CSRm were injected into CSRe. The revolution periods of the stored ions, which depend on the mass-to-charge ratios of the stored ions, were measured with a time-of-flight detector system. The results show that the mass resolution around 8×10-6 for Δm/m is achieved.