Wobbling Magnet Design for Beam Line of CYCIAE-30 Medical Cyclotron
- Received Date: 2006-07-06
- Accepted Date: 2006-08-25
- Available Online: 2007-03-05
Abstract: The beam extracted from cyclotron is of Gaussian distribution, but in many applications the uniformly distributed beam is needed. There are some methods to achieve uniform beam, but the uniformity is less than expected. This paper introduces the process of the wobbling magnet design which can get the high uniform beam. On the base of the design theory and the measurement to the wobbling magnet, we can confirm that there has high uniform beam on the target under the wobbling magnet effect. It is the first time to design and manufacture the wobbling magnet by ourselves in our country, and the measurement results show that the magnetic field in the wobbling magnet are satisfied the requirement of the storage wobbling magnet.





Abstract
HTML
Reference
Related
PDF












DownLoad: